Abstract

Detection of air bubbles in the blood is important for various medical treatments that use Extracorporeal Blood Circuits (ECBC), such as hemodialysis, hemofiltration and cardio-pulmonary bypass. Therefore a reliable air bubble detector is needed. This study presents the design of a new fault tolerant air bubble detector. Triple Modular Redundancy (TMR) method is used on the sensor section. A voter circuit of the Triple Modular Redundancy will choose one of three sensor outputs to be processed further. Application of Triple Modular Redundancy will prevent errors in the detection of air bubbles, especially if the sensor fails to work.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.