Abstract

For a variety of applications, such as in miniaturized machines, tools for minimal invasive surgery, or coronary stents, microscale components are used. For all these components, their dimensions are far below the size of conventional test specimens, and thus the grain size can approach the dimension of the cross section in these microscale components. According to experimental results, large differences in the mechanical behavior of the material occur between single- and polycrystalline test specimens. Therefore, oligo-crystalline microstructures are defined as a transition between single- and polycrystal. To investigate and understand the fundamental impact of oligo-crystalline microstructures on the mechanical behavior of the material, thin wires made of the austenitic CrNiMo steel 316LVM were fatigued. The choice of the material is justified, because it is one of the most frequently-used materials for coronary stents, and only a small amount of research has been done on oligo-crystalline microstructures of this material. Solution were annealed and 10% cold drawn oligo-crystalline wires were compared. The cold drawn wires exhibit an endurance limit of 450 MPa, which is significantly higher compared to solution annealed oligo-crystalline wires (250 MPa). Electron backscattering diffraction (EBSD) measurements of the fatigued wires show massive grain rotations, which lead to orientation changes within the grains. Sometimes, the deformation of a whole structure is concentrated on just one or only very few grains, with a particularly high Schmid factor (>0.44).

Highlights

  • The CrNiMo-steel type 316LVM (DIN EN 1.4441) is one of the most used metals for coronary stents [1,2]

  • Electron backscattering diffraction (EBSD) measurements of the fatigued wires show massive grain rotations, which lead to orientation changes within the grains

  • The cold drawing of the wires increased the tensile strength from an average of 415 MPa to 804 MPa and decreased the increased the strain tensile from strength an average of 415

Read more

Summary

Introduction

The CrNiMo-steel type 316LVM (DIN EN 1.4441) is one of the most used metals for coronary stents [1,2]. Its austenitic microstructure provides good properties for ductility and strength [3,4]. Clinical studies revealed a significantly lower risk for early and late in-stent restenosis for thin-strut stents compared to stents with thick struts [1,5,6,7,8]. In addition to clinical studies, up to now, only a small amount of research has been done on the mechanical properties of oligo-crystalline components. Corresponding computer simulations reveal that the properties of these oligo-crystals can neither be considered as single-, nor as poly, crystalline behavior [9,10]. In previous studies, authors have shown strongly deviating monotone mechanical properties of thin wires, depending on the local crystallographic grain orientation [11,12]

Methods
Results
Conclusion
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.