Abstract

This study focuses on the fatigue behavior of titanium dental implants as-received, with a grit-blasted surface and with a new bioactive surface treatment (2Steps). The 2Step process consists of (1) an initial grit-blasting process to produce a micro-rough surface, followed by (2) a combined thermo-chemical treatment that produces a potentially bioactive surface, that is, that can form an apatitic layer when exposed to biomimetic conditions in vitro. The 2Step treatment produced micro-rough and apatitic coating implants. Residual stresses were determined by means of X-ray diffraction. The fatigue tests were carried out at 37°C on 500 dental implants, and the S-N curve was determined. The fatigue-crack nucleation for the different treatments was analyzed. The fatigue tests show that the grit-blasting process improves the fatigue life. This is a consequence of the layer of compressive residual stresses that the treatment generates in titanium surfaces. Dental implants that had its surfaced prepared with the 2Step procedure (grit-blasting and thermo-chemical treatment) had its fatigue life decreased by 10% due to the incorporation of oxygen to the surface and the relaxation of the compressive residual stress produced by the heat treatment. Thermo-chemical treatment is an excellent compromise between the improvement of bioactive and mechanical long-life behaviors.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.