Fatigue crack growth in orthopaedic locking compression plates (LCP), under four-point bending, has been simulated by the extended finite element method (xFEM) to assess LCP optimal geometry in respect to remaining life. Loads corresponded to those occurring in human tibia during gait cycle for different body weights (60, 90 and 120 kg). Experimental investigation consisted of tensile testing and fatigue crack growth rate testing of Ti-6Al-4 V alloy, as used for LCP, by standard testing using tensile testing machine and Fractomat, respectively. Results showed that the remaining life of LCPs strongly depend on plate geometry and patient body weight (load).

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.