Abstract

A previous method for determination of the radiochemical purity (RCP) value of [125I]sodium iothalamate uses two paper strips and solvents (total developing time is approximately 2.5 h). To simplify and shorten the RCP testing procedure, our laboratory has developed a single-strip chromatography method that not only distributes free 125I and [125I]sodium iothalamate to different relative front (Rf) locations, but is also faster and easier to perform. RCP of [125I]sodium iothalamate was determined with the use of a 10-cm instant thin-layer chromatography strip impregnated with polysilicic acid gel (ITLC-SA) as the solid phase, and a mobile phase of 2-butanol:acetic acid:water (140:2.5:70, v/v). By using autoradiography and counting the strip segments in a gamma counter, our results indicated that free 125I migrated to Rf = 0.89-1.00 while the [125I]sodium iothalamate moved to Rf = 0.44-0.67. The total developing time for the single-strip ITLC-SA system was approximately 1 h.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.