Abstract

To evaluate the diagnostic efficacy of fast T2-weighted magnetic resonance (MR) imaging sequences on image quality, hepatic lesion detection, and lesion conspicuity. Three breath-hold, fast T2-weighted sequences with turbo-spin-echo (TSE), half-Fourier acquisition single-shot TSE (HASTE), and inversion recovery (IR) HASTE techniques were examined for 43 lesions in 20 consecutive patients. Evaluation was performed qualitatively on image quality and lesion detectability and quantitatively on lesion conspicuity by using lesion/liver signal-intensity and contrast-to-noise ratios. Artifacts were significantly less present on the HASTE sequence (p < 0.01). Both TSE and HASTE sequences detected 39 lesions (91% each); the IR HASTE sequence detected 37 (86%). IR HASTE sequence showed a significantly higher signal-intensity ratio than did the others (p < 0.01). Breath-hold TSE versus breath-hold HASTE or IR HASTE is still the most robust sequence in lesion detection, image quality, and lesion conspicuity. However, the HASTE sequence offers good lesion detection and image quality, and the IR HASTE has a better signal-intensity ratio.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call