Batch injection analysis (BIA) with amperometric detection, using a pyrolytic graphite electrode modified with cobalt phthalocyanine (PG/CoPc), was employed for determination of α-lipoic acid (ALA) in pharmaceutical product and in synthetic urine samples. The proposed BIA method is based on the application of a potential of +0.9V vs. Ag/AgCl, KCl sat, enabling quantification of ALA over a concentration range from 1.3×10−6 to 1.0×10−4molL−1, with a detection limit of 1.5×10−8molL−1. A sampling rate of 180 injections per hour was attained and measurements of the reproducibility of successive injections (100µmolL−1 ALA on the same electrode) showed a RSD of 2.11% for 40 successive injections. The new sensor was utilised for ALA quantification in a dietary pharmaceutical supplement and in synthetic urine and the results obtained for both samples were compared with parallel analysis using high performance liquid chromatography (HPLC), the method recommended by the United States Pharmacopeia. The results obtained were similar (at a 95% confidence level) and in the case of the synthetic urine sample (prepared with a known amount of ALA) the recovery was situated between 98.0% and 102.6%.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.