Abstract

Abstract Some years ago a fast neutron spectrometer method was developed based on the etched track density determinations of elastically scattered protons and using varying radiator thickness at fixed etching time and varying etching times at fixed radiator thicknesses. Some new results in this field are presented.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.