Abstract

This paper reports on the performance of the inorganic scintillator caesium hafnium chloride (CHC) under exposure to the mixed radiation field of an AmBe neutron source and coupled to a silicon photomultiplier (SiPM). The neutron response is determined using the pulse shape discrimination charge comparison technique which can clearly identify both the (n,α) and (c) reactions in the material. Figures of merit for the pulse shape discrimination are presented and the quenching of the different channels is assessed through comparison to Monte Carlo simulations.