Abstract

EJ-276 is a recent development in plastic scintillators that comes with improved pulse shape discrimination over its predecessor EJ-299-33. In this work, the neutron response functions, proton light output function, and detection efficiency of EJ-276 were measured using mono-energetic neutron reference fields at the national metrology laboratory, Physikalisch-Technische Bundesanstalt, in the neutron energy range 1.5–19 MeV. In particular, the proton light output in the range 14–19 MeV is presented for the first time. The optimization of Pulse Shape Discrimination has also been investigated using an 241Am-Be source and results are presented. Finally, we study how, using a shadow-cone method and time of flight method, we can discriminate scattered and other spurious neutrons from the mono-energetic beam. The characterized EJ-276 detector is to be used in the monitor of the neutron fields and gamma background at the new Neutron Irradiation Laboratory for Electronics facility at the Rutherford Appleton Laboratory.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.