Abstract

Fast neutron imaging was performed using a beamline of the 10MW research reactor of the Budapest Neutron Centre, Hungary. A simple, low-cost 2D area detector has been used featuring a 8mm thick BC400 plastic scintillator converter screen and a CCD camera. A spatial resolution of around 1.3mm has been achieved. Typically 10min long exposures were needed to obtain reasonable quality radiographic images. For tomographic imaging typically several hours of acquisition were needed to obtain reasonable quality on non-symmetric and larger (e.g. 10×10×10cm3) objects. Due to the presence of a significant gamma background at the experimental position, massive (30cm thick) lead shielding and filtering was applied to the beam. The gamma contribution was mostly baseline independent of the object imaged and therefore could be subtracted, whereas the direct gamma contribution from the beam to the imaging detector signal is estimated to be less than 1%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.