Abstract

A method is presented for the rapid acquisition of temperature maps derived from phase difference maps. The temperature-dependent chemical shift coefficients (TDCSCs) of various concentrations of aqueous cobalt and dysprosium-based compounds were measured. The largest TDCSC calculated was for 100 mM DyEDTA, which had a TDCSC of -0.09 PPM/K; 160 mM CoCl2 had a TDCSC of -0.04 PPM/K. These temperature-dependent chemical shifts (TDCSs) result in phase changes in the MR signal with changing temperature. Agarose phantoms were constructed with each paramagnetic metal. A fast gradient-echo (FGRE) MR image was acquired to serve as the baseline image. A "test" MRI procedure was then performed on the phantom. Immediately afterwards, a second FGRE MR image was acquired, serving as the probing image. Proper image processing as a phase difference map between the probing image and the baseline image resulted in an image which quantitatively described the temperature increase of the phantom in response to a particular "test" imaging experiment. Applications of this technique in assessing the safety of pulse sequences and MR coils are discussed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.