Abstract

The present paper reports the development and the application of a microplate assay for the fast and individual determination of thiols and sulfides. The method is based on the reaction of ethyl propiolate with thiols and sulfides, yielding thioacrylates absorbing in the UV range. Thioacrylate generated by derivation of sulfide with EP presents a bathochromic shift of maximal absorbance wavelength of 16nm compared to thioacrylates formed with thiols. EP concentration, reaction time and derivatization pH were carefully optimized in order to separately quantify thiols and sulfides using a spectral deconvolution procedure in 15min. The detection limits were 0.6 and 2.8μM (63.7 and 92.4μg·L−1) for mercaptopropionic acid (used as reference thiol compound for calibration) and dissolved hydrogen sulfide respectively. Non-thiol sulfur compounds (methionine, sulfates and sulfites) or other nucleophile compound did not display any interference in the determination of MPA or sulfides. Interference from metals was suppressed with application of a reduction step with TCEP and DTPA. Finally, the method has been successfully applied to the analysis of raw and treated wastewater.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.