Abstract

Time counting optimization in gamma-ray spectrometry was established using four HPGe detectors and four counting geometries. Deionized water samples with a mixed radioactive solution of 133Ba, 134Cs, 106Ru, 137Cs, 65Zn, and 60Co were prepared in volume from 0.1 to 3 L, and measured three times, for all detection systems, and for seven different counting times, ranging from 1,000 to 150,000 s. The obtained results were compared to the reference values from solutions with well-known activity concentrations samples, for minimum detectable amount (MDA) calculation. A counting time of 50,000 s was found to be generally sufficient to reach agreement between the preset and actual counting times.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.