We have implemented a new magnetization transfer (MT) technique, to decrease total imaging time and thus improve efficiency of MT based MRI. By judiciously rescaling different RF pulses, and utilizing maximum transmitter gain, we were able to generate very short and powerful MT pulse(s). MRI and MRA in normal and patients with MS were performed, with and without MT application. Phantom experiments indicated 1-2 kHz offset to be the optimal. MT based MRA demonstrated substantial improvement in conspicuity of small vessels with minimum increase in imaging time. MT based MRI showed 30% improvement in MT contrast as well. Importantly, both peak and average SAR values were within FDA approved safety limits. >

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.