Abstract

We present a setup for fast, low-bias (≤1mV) DC transport measurements with μs time resolution for high ohmic resistance (≈20kΩ) semiconducting samples. We discuss the circuitry and instrumentation for the measurement approach that can be applied to any kind of semiconductor device or (gated) two-dimensional material and demonstrate the main measurement artifacts in typical measurements by means of circuit simulation. Based on the latter, we present a simple two-step protocol for eliminating the measurement artifacts reliably. We demonstrate the technique by measuring the transitions between quantum Hall plateaus in the HgTe quantum wells and resolve plateaus as short-lived as 100 μs.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call