Abstract
Kinetic resolution of (R,S)-atenolol is a faster strategy to produce (S)-atenolol. Since this racemate is a less soluble compound, resolution of its ester offers high concentrations in the process. A good analytical method is required to observe the enantiomer concentrations. This paper described application of ultra-fast liquid chromatography on the atenolol ester separation using different resolution media and analytical procedures. Chiralcel OD column resolved the ester. The chromatograms indicated different characteristics of the process. The enantiomers could be recognized by the column in less than 1 (one) hour. Symmetrical peaks were obtained, but several procedures produced peaks with wide bases and slanted baselines. Efficient enantioresolution was obtained at high mobile phase flow rate, decreased concentration of amine-type modifier, but increased alcohol content in the mobile phase. High UV detection wavelength was required. At 1.0mL/min, the (90/10/0.5) composition resulted α=1.46 and RS =0.9998 that were good separation.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.