Abstract

Optical computed tomography (optical-CT) of 3D radiation dosimeters is a promising avenue for delivering an economic and reliable quality control of radiotherapy treatments such as intensity modulated radiotherapy, brachytherapy and stereotactic radiosurgery. The main problems in transferring 3D dosimeters to clinical setting have been in (1) the complexity of manufacture and behaviour of 3D dosimeters and (2) time-consuming readout and analysis of 3D dosimeters. This paper addresses the readout problem by showing that fast (20 min tomography scan), precise (projection absorbance signal-to-noise ratio is greater than 100:1 across the absorbance range 0.2 to 1.5) and accurate (good linearity in the calibration curve) measurements are possible using a novel method of optically scanning a laser beam across the 3D dosimeter.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.