Abstract

An improved flow injection cytometry (FIC) system suitable for fast cellular kinetic measurements and on-line dilution is described. The instrument allows for measurements from 1.2 s (+/- 0.05 s) after the initiation of mixing, up to any time thereafter. Crucial factors in determining fast kinetic measurements, such as the displacement volume for sample introduction, rate of transport to cytometric interrogation point, and the mixing speed are evaluated and discussed. By varying the volume aspirated, this instrument can facilitate the dilution of cells and/or reagent over a range of one order of magnitude. The fast kinetic and on-line dilution capabilities were demonstrated by on-line staining of DNA in trout erythrocytes with 4',6-diamidino-2-phenylindole (DAPI). The utility of the instrument for measurement of enzyme kinetics was illustrated using human lymphocytes, measuring the glutathione S-transferase (GST) catalyzed reaction between monochlorobimane (MCB) and glutathione (GSH).

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.