Abstract

Confocal Raman microscopy (CRM) is extensively employed in numerous fields because it provides molecular fingerprint information without destroying or labeling samples. For samples with large topographic fluctuations, existing technology requires axial scanning to achieve axial focusing, which further increases the time required for Raman mapping. Herein, a method called non-axial scanning confocal Raman microscopy (NASCRM), which constructs an axial focusing system with a large range and high precision by modifying the wide-field imaging system, is proposed. The axial focusing system combined with CRM can simultaneously realize in situ high-precision topography imaging and Raman mapping. Theoretical analyses and preliminary experiments reveal that the axial focusing resolution of NASCRM is approximately 3 nm and the sensing range is approximately 60 μm. Compared with CRM using confocal microscopy (CM) focusing (CM-CRM), the speeds within sensing range for axial focusing and Raman mapping are improved by 5.3 and 2.9 times, respectively. This study provides a novel method for both conventional CRM and CRM-based imaging in various fields.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.