Abstract

A fast infrared array spectrometer (FIAS) with a thermoelectrically cooled 160-element PbSe detector was demonstrated using measurements of instantaneous infrared radiation intensities simultaneously over the 1.8–4.9 μm wavelength range at a sampling rate of 390 Hz. A three-point second-degree Lagrange interpolation polynomial was constructed to calibrate the FIAS because of the nonlinear response of the infrared array detector to the incident radiation beam. This calibration method gave excellent measurements of blackbody radiation spectra except for a narrow band at wavelength of 4.3 μm due to absorption by room carbon dioxide, which is one of the two major gas radiation peaks (2.7 and 4.3 μm) from the lean premixed hydrocarbon/air combustion products in the midinfrared spectrum. Therefore, the absorption coefficient of room carbon dioxide was conveniently measured on site with the blackbody reference source, and was used in the calibration of the FIAS and also in the calculations of the radiation spectra. Blackbody tests showed that this procedure was effective in correcting for the room carbon dioxide absorption in the radiation spectra measured by the FIAS. For an example of its application, the calibrated FIAS was used to measure spectral radiation intensities from three lean premixed laminar flames and a premixed turbulent jet flame for which reference data with a grating spectrometer were available for comparison. The agreement between the FIAS measurements and the reference data was excellent.