Abstract

The Ångström method is a promising thermal diffusivity measurement method for microfibers. Based on the heat-loss Ångström method and micro-Ångström method, the FFT (Fast Fourier Transform) Ångström method can be used to shorten the testing time and enhance the testing robustness of thermal conductivity. The FFT Ångström uses multi-frequency superimposed input signals and then uses the FFT for frequency division analysis. Since different frequencies are input at the same time, the measuring time is greatly reduced. Moreover, because different frequencies are input in the same environment, the random and environmental perturbations of the frequencies are the same, enhancing the fitting robustness. In contrast, the single feeding strategy is relatively time-consuming, and its measurement homogeneity for different frequencies cannot be guaranteed. By comparing the measurement results from a multi-frequency input and separated single-frequency input, the FFT shows good feasibility and robustness. It may also have great potential in other types of thermal wave measurements.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.