Abstract

A novel fluid mixing strategy was developed which significantly enhanced the efficiency of DNA hybridization. A pneumatic micro-mixing device consisting of two pneumatic chambers and an underneath DNA microarray chamber was built up. The fluid in the array chamber was pneumatically pumped alternately by the two pneumatic chambers. The chaotic oscillatory flow caused by the pumping greatly intensified the fluidic mixing. A homogeneous distribution of the tracer dye solution in the microarray chamber was observed after 2s mixing with a pumping frequency of 24Hz. Microarray DNA hybridization was substantially accelerated using this device, and the fluorescence intensity showed a plateau after oscillating 30s at room temperature. The corresponding signal level of the dynamic hybridization was 12.5-fold higher than that of the static hybridization performed at 42°C. A signal-to-noise ratio of 117 was achieved and the nonspecific adsorption of the targets to the sample array was minimized, which might be attributed to the strong shearing force generated during the pneumatic mixing process.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.