Abstract

X-ray fluorescence (XRF) analysis was successfully applied for the determination of Ca, Cu, Fe, Mn, Ni, Pb, V and Zn on aerosol filters with a diameter of 150 mm. Ambient aerosol was collected on quartz-fibre filters by Digitel high-volume samplers. For XRF analysis three small filter pieces with a diameter of 32 mm were cut out of each filter, considering the radial dependence of elemental distribution on the filter. Elemental concentration decreases by 8-10% from the centre to the outer circle for coarse mode elements. Distribution of fine mode elements display only small radial dependence (2-4%). Results of quantitative analysis by XRF were compared with atomic spectrometric methods (GF-AAS, ICP-OES). The good agreement between the methods confirmed the application of XRF spectrometry for routine analysis of filter samples loaded with ambient aerosol.