Abstract

A fast and robust analytical method for amperometric determination of hydrogen peroxide (H 2O 2) based on batch injection analysis (BIA) on an array of gold microelectrodes modified with platinum is proposed. The gold microelectrode array ( n = 14) was obtained from electronic chips developed for surface mounted device technology (SMD), whose size offers advantages to adapt them in batch cells. The effect of the dispensing rate, volume injected, distance between the platinum microelectrodes and the pipette tip, as well as the volume of solution in the cell on the analytical response were evaluated. The method allows the H 2O 2 amperometric determination in the concentration range from 0.8 μmol L −1 to 100 μmol L −1. The analytical frequency can attain 300 determinations per hour and the detection limit was estimated in 0.34 μmol L −1 (3 σ). The anodic current peaks obtained after a series of 23 successive injections of 50 μL of 25 μmol L −1 H 2O 2 showed an RSD < 0.9%. To ensure the good selectivity to detect H 2O 2, its determination was performed in a differential mode, with selective destruction of the H 2O 2 with catalase in 10 mmol L −1 phosphate buffer solution. Practical application of the analytical procedure involved H 2O 2 determination in rainwater of São Paulo City. A comparison of the results obtained by the proposed amperometric method with another one which combines flow injection analysis (FIA) with spectrophotometric detection showed good agreement.