Abstract

We present a fast, automatic method for mitral annulus excursion measurement using pocket-sized ultrasound (PSU). The motivation is to provide PSU users with a rapid measurement of cardiac systolic function. The algorithm combines low-frame-rate tolerance, computational efficiency and automation in a novel way. The method uses a speckle-tracking scheme, initialized and constrained by a deformable model. A feasibility study using 30 apical four-chamber PSU recordings from an unselected patient population revealed an error of (mean ± SD) –1.80 ± 1.96 mm, p < 0.001, when compared with manual anatomic m-mode analysis using laptop scanner data. When only septal side excursion was measured, the mean error was –0.27 ± 1.89 mm, p < 0.001. The accuracy is comparable with previously reported results using semiautomatic methods and full-size scanners. The computation time of 3.7 ms/frame on a laptop computer suggests that a real-time implementation on a PSU device is feasible. (E-mail: sten.r.snare@ntnu.no)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.