Abstract

Apolipoprotein monitoring is useful for diagnosing cardiovascular diseases, as they are risk factors of arteriosclerosis and other neutral fat-related diseases. Liquid chromatography-tandem mass spectrometry (LC-MS/MS) is advantageous for simultaneous apolipoprotein quantification, differentiation, and standardization including their isoforms. However, fast and straightforward sample preparation that retains quantification accuracy remains challenging in clinical MS. We developed a simultaneous assay for serum apolipoprotein A-I (ApoA-I), apolipoprotein B100 family, and apolipoprotein C-III (ApoC-III) using a high-throughput LC-MS/MS platform coupled with a BRAVO system. The assay was simplified by using sodium deoxycholate and trypsin/lys-C without reduction and alkylation steps. Simple sample preparation reduced turnaround time by 1.5h and neat goat serum was chosen as an optimal calibration matrix for accurate protein quantification. Assay precision, linearity, correlation, accuracy, limit of detection (LOD), limit of quantitation (LOQ), and carryover were validated according to CLSI guidelines over 41days using more than 100 human serum samples. Good correlation compared with turbidimetric immunoassay (TIA) was observed by Deming regression for all analytes. A high-throughput LC-MS/MS and BRAVO assay for simultaneous apolipoprotein analysis was validated using a simple preparation method with a human serum calibrator in goat serum matrix. The assay is readily expandable to include other target serum proteins and/or their isoforms.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call