Abstract

We have developed an HTS SQUID system for both time and frequency domain measurements of magnetic signals from the Brownian relaxation of liquid-suspended magnetic nanoparticles (MNPs) coated with antibodies. A planar YBa <sub xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">2</sub> Cu <sub xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">3</sub> O <sub xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">7-x</sub> dc SQUID gradiometer with a baseline of 3 mm allows stable operation of the measurement system in an unshielded environment. Agglomeration of MNPs induced by prostate specific antigen-antibody binding complexes is characterized with our system and benchmarked with frequency domain measurements performed by Imego AB using an induction coil magnetometer. We estimate an upper bound for our immunoassay analyte sensitivity of 1.8 ng/radicHz.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.