Abstract

To establish and evaluate the analytical and clinical performance of the Flash20 SARS-CoV-2 nucleic acid rapid detection system free of RNA extraction. The limit of detection (LoD) was determined using a negative nasopharyngeal swab matrix spiked with different concentrations of SARS-CoV-2 virus; a total of 734,337 reference sequences of viral genomes from GenBank were used for the in-silico analysis to assess the inclusivity of the assay. The specificity of the system was evaluated by testing 27 medically relevant organisms. A total of 115 clinical specimens were collected and tested on the Flash20 SARS-CoV-2 detection system and with an FDA-approved comparator test to assess the clinical performance of the system. The LoD of the Flash20 SARS-CoV-2 detection system is 250 copies/mL with a positive rate ≥90% (n = 20); alignments results showed that over 99% identity of the primer and probe of the Flash20 SARS-CoV-2 nucleic acid rapid detection system to the available SARS-CoV-2 sequences; the omicron samples tested 100% positive. None of the 27 organisms showed cross-reactivity with the Flash20 SARS-CoV-2 nucleic acid rapid detection system. Among all the 215 clinical samples, the Flash20 SARS-CoV-2 nucleic acid rapid detection system exhibits a high sensitivity of 99.24% (131/132) and 100% (83/83) specificity. The nucleic acid rapid detection system provides sensitive and accurate detection of SARS-CoV-2 free of RNA extraction. The high sensitivity and short time to results of approximately 35 minutes may impact earlier infection control and disease management.