Abstract

Saxitoxin (STX) and neosaxitoxin (NEO) are water-soluble toxins and their cleanup in bio-matrix is a hot topic but difficult problem. A fast and quantitative determination method for STX and NEO in urine was developed using ultra performance liquid chromatography-triple quadrupole mass spectrometry (LC-MS/MS) based on the cleanup of solid phase extraction (SPE) with hydrophilic interaction (HILIC) mechanism. Acetonitrile/methanol/water mixture was used to extract the toxins in urine. Polyamide (PA) was used as HILIC SPE material to clean the toxins in sample matrix. The limits of detection were 0.2ngmL-1 for STX and 1ngmL-1 for NEO in urine. The linear ranges were 0.5ngmL-1-99.2ngmL-1 with the correlation coefficient of r=0.9992 for STX and 2.1ngmL-1-207ngmL-1 with r=0.997 for NEO in urine matrix. The recoveries at three spiking levels were 81.5%-117% with the relative standard deviations (RSDs) of 5.4%-8.5% for STX and 89.0%-118% with the RSDs of 6.7%-9.1% for NEO. STX was found in all the 6 patients' urines while NEO was only found in one sample from an intoxication case.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.