Abstract

A new type of high-performance liquid chromatography (HPLC)–size exclusion chromatography (SEC) system with ultraviolet (UV) absorbance detection and non-dispersive infrared (NDIR) detection of total organic carbon is described. The introduction of an online degassing tube and a low-volume HPLC column helped to reduce the analytical time and increase the sensitivity of the SEC system. This study is the first in which linear calibration curves ( R 2 > 0.99) were obtained for both UV absorbance and NDIR data for polystyrene sulfonate standards, which are the most suitable standards for molecular size analysis of aquatic humic substances as well as dissolved organic matter (DOM). Using the calibration curves, the molecular size distribution of DOM in water collected from Lake Kasumigaura and in pore water from lake sediments was estimated. Most of the DOM had a molecular weight less than 4000 Daltons (Da), and the amount of low-molecular-weight DOM (∼2000 Da) with low UV absorbance increased with depth in the sediment pore water. This result shows the importance of combining quantitative analysis by NDIR detection with qualitative analysis by UV detection to determine the chemical and physical properties of DOM. The possible sources and reactivity of DOM in Lake Kasumigaura and its sediment pore water are also discussed.