Abstract

The fractionation of positional isomers of a PEGylated protein is quite challenging as these have similar molecular weight, and only very slightly different surface charge. In this study, cation exchange z2 laterally-fed membrane chromatography (z2LFMC), which has been shown to be suitable for high-speed, high-resolution protein purification, was used to fractionate positional isomers of mono-PEGylated lysozyme. The performance of the z2LFMC device was compared with a commercial preparative cation exchange column having the same volume and ligand. PEGylated lysozyme purification experiments showed that while the positional isomers of mono-PEGylated lysozyme could not be satisfactorily resolved using the preparative commercial cation exchange column, almost baseline resolution of these could be achieved using the z2LFMC device. Moreover, the z2LFMC device-based process was faster by an order of magnitude. The results discussed in this paper demonstrate that z2LFMC is a superior alternative to column-based chromatography for challenging protein separations, such as the one discussed here, both in terms of speed and resolution.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.