Abstract
An analytical method for simultaneous determination of total polyphenol content (TPC) and antioxidant activity (AA) of wines (white and red wines) and oenological tannins, using a flow injection system with sequential diode array and electrochemical amperometry detectors (DAD-ECD), was proposed. The signal at 280 nm provided aggregate data for TPC. The anodic peak related to wine phenolic oxidation was scanned using pulsed integrated amperometry over the potential of 800 mV vs. Ag/AgCl, to obtain AA. Serial dilutions avoided the poisoning at the glassy carbon (GC) electrode and the linear response obtained with both detectors was compared with spectrophotometric assays commonly used in oenology laboratory. Intraday and interday analytical repetitions showed a good repeatability and reproducibility (relative standard deviation RSD < 6% for both detectors), and the satisfactory relationship between the proposed coupled flow injection/DAD-ECD and the classic UV methods (R2TPC = 0.9967; R2DPPH = 0.9621) confirmed the efficacy of flow injection analysis with a coupled detection system, for the reliable quality control of wine and wine-related products.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.