Abstract

Amplitude modulation (AM) scanning force microscopy (SFM) is superior to frequency modulation SFM in simplicity, sensitivity, and stability, but is still replaced by the latter because it is too slow when the Q factor is high (bandwidth < 0.5 Hz for Q > 50,000 and resonant frequency ω0 < 50 kHz). We report a close-loop AM detector that has an 18 Hz bandwidth, better than 1 mHz frequency resolution and excellent response to step frequency changes even for Q ∼ 60,000 and ω0 ∼ 32 kHz. Its superiority is well shown by the comparison of magnetic force microscope images taken under the new and old AM detection modes with the tip and scan area (videotape sample) being unchanged. Also important is that shifting the driving frequency from near the resonance peak to further away from the peak does not decrease the frequency resolution as much as we expect (but can increase the response speed).

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.