Abstract
In this paper, an algorithm for detection of reference spheres from digital panoramic radiographic images is presented. The proposed algorithm was tested on a database of 107 digital panoramic radiographic images which were used for dental diagnostics. Results show that the proposed method exhibits for detection of reference spheres, a sensitivity of 97.33% and specificity of 93.85%. Performance time differed between 0.55 and 2.36 s depending on image size. The aim of this work was to provide a fast ellipse detection algorithm to reduce measuring time on preoperative implant planning by lowering the computational cost.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.