Abstract
All free-sulfhydryl compounds tested produced false-positive reactions in the Legal test for ketones. The color developed in the ketone pad of urine dipsticks [N-Multistix SG, Multistix 10 SG (Ames), and Chemstrip 9 (Boehringer-Mannheim)] was misinterpreted for ketone bodies, both by visual and automated reading. In contrast to the reaction with true ketones, a drop of glacial acetic acid added onto the ketone pad of dipsticks discharged the false-positive red color. A red-violet also developed instantly with free -SH compounds in the Acetest tablet assay (Ames), but quickly faded. In general, the presence of acidic groups such as -COOH and -SO3H in the structure appeared to increase the nitroprusside reactivity of free -SH compounds, whereas the presence of a -NH2 group appeared to decrease it. Currently, false-positive ketone reactions ascribable to a free -SH group are most likely to be seen for urine containing mesna. The false-positive test for ketones caused by free -SH compounds can be recognized and ruled out by proper procedures. On the other hand, this chromogenic reaction with free thiols might be used for monitoring urinary excretion of mesna.
Published Version (Free)
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.