Abstract
The 3f Aortic Bioprosthesis (Medtronic, Inc, Minneapolis, MN) is a stentless aortic valve with a novel design that resembles a "tube within a tube." Although it has the potential for improved durability and hemodynamic performance, long-term data on this valve remain elusive. We present here 3 patients in whom postoperative echocardiography revealed significantly elevated transvalvular gradients of the 3f valve while transcatheter gradients proved to be negligible. By virtue of the unique design of the 3f bioprosthesis, great caution should be taken when interpreting echocardiographically derived gradients.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.