Abstract

Isolation of Mycobacterium avium-intracellulare from a number of specimens cultured with the Bactec 460 TB system was suspected to be due to carry-over contamination. The analysis of culture records and of the typing results confirmed this hypothesis. The problem seems to have been eliminated after replacement of the needle heater and increase of the needle temperature. We recommend that a number of precautions should be taken to reduce the risk of reporting false positive mycobacterial cultures.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.