Abstract

Rotating hinge knee prostheses are known to provide inherent stability. Yoke fractures of the hinged tibial insert of modern generation rotating hinge devices are a matter of continued concern. The aim of this study was to describe incidence and management of yoke fracture of the LPS™ hinged tibial insert. Retrospective data analysis of two institutions identified 40 patients with a LPS™ total knee arthroplasty. Implant survival and prosthetic complications was calculated according to Kaplan-Meier. Out of the group of 40 patients, four fractures of the metal yoke occurred in four cases (failure rate: 10%). Furthermore, a second fracture occurred in two patients. The overall revision-free prosthetic survival was 57% at 38 months, while prosthetic survival until yoke fracture was 86% at 38 months. Handling yoke fractures as mechanical complication includes replacing the hinged insert, stabilization of the joint and joint line height preservation in order to decrease the cantilever effect at the insert-base plate interface.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.