Abstract
The most commonly reported failure mode of sliding hip screws in published literature is cut-out of the lag screw. This study investigates the resistance to failure of the femoral head, with lag screws used in two types of sliding hip screws, the gamma locking nail (Howmedica) and the dynamic hip screw (DHS) (Synthes). The investigation consisted of biomechanical tests under static loading conditions on 12 pairs of cadaveric femoral heads, to establish the failure loads due to screw cut-out for the two implant lag screws. The gamma nail appeared to reduce the tendency to cut-out in the osteoporotic bone (soft) associated with elderly patients in whom these devices are commonly used (p<0.05). In high density bone (hard) the gamma lag screw also appeared to be stronger, because the DHS showed a tendency to bend. The larger diameter of the gamma nail lag screw resists bending and appears to reduce the risk of cut-out compared with the DHS.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.