Abstract
Fractures resulting from wear and fatigue process have been identified as the main causes of failure in biomaterials, especially in implants that act in place of bone or other hard tissue, as they are subject to conditions involving severe cyclic loadings. In biomaterialscase, the types of failures mentioned above must also be evaluated under the effect of degradation or corrosion, due to the direct contact with body fluids. The present research analyzed the fatigue induced by corrosion fracture of an orthopaedic implant for total knee replacementproduced with an austenitic ASTM F138 stainless steel. The morphology, compositions of the interfaces and subsequent corrosive pitting were characterized by stereoscopy and scanning electron microscopy (SEM) coupled with energy dispersive spectroscopy (EDS). Stress concentration, inclusions and high carbon levels were the main reasons of failure.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.