Abstract

In view of conflicting results on the fading characteristics of lithium fluoride, a careful reinvestigation has been made on possible fading of the commonly used phosphor, lithium fluoride TLD-100. Measurements over a twelve week period on freshly irradiated powder from a new batch, and also on some old powder, indicate fading of 5 ± 3%, most of which may be due to changes in sensitivity after annealing. Recommendations are made for the proper use of the phosphor in radiation dosimetry applications.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.