Abstract
Speech audibility may be defined as that proportion of a speech spectrum which is above a person's threshold. To optimize speech audibility with a hearing aid, several measures are needed. These include quantification of a speech spectrum, measures of hearing sensitivity, and measures of the "real ear" gain of the hearing aid. Some procedural factors must be considered to obtain adequate measures in a typical clinical setting. Those considered here are: (1) a modified Articulation Index to quantify a speech spectrum, (2) specification of hearing sensitivity with a narrowband sound field reference where the out-of-band rejection rate of the sound field stimulus is greater than twice the slope of the hearing loss, and (3) use of functional gain (measured directly or estimated using earphone and sound field results provided that the sound field stimulus has the required characteristics for measuring hearing sensitivity in the sound field) as a measure of the real ear gain of the hearing aid. Guidelines are given for the practical measurement of speech audibility in a typical clinical setting. The guidelines are appropriate for all measures of real ear gain including those obtained with all probe tube systems.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.