The treatment of Aortic regurgitation using TAVR has not met with excellent outcomes that TAVR did with Aortic stenosis treatment due to differences in patient anatomies, and manifestation of the disease at the native valve annulus. With exception of JenaValve device which relies on proprietary clipping mechanism to anchor the valve, off-label use of TAVR devices that are otherwise indicated for use only to treat AS disease, relied on valve-in-valve for anchoring to prevent paravalvular leakage and valve embolization. The under-expansion of the inside valve due to stiffness and geometry of the outside valve has an impact on leaflet coaptation leading to additional stresses on the leaflets and decreased durability. This article explores the stent frame and leaflet design considerations and their impact on valve durability in AR patient anatomies that rely predominantly on valve-in-valve configuration for anchoring.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.