Abstract

AbstractA series of quaternary ammonium salts was analysed by the ionspray technique under different conditions in order to study the influence of some important factors on the ionization efficiency, The factors taken into account were droplet charging, sample ion, counter ion, solovent composition, and concentration. Measurement of the spray current leaving the nebulizer has shown that the spray current increases linearly with sample concentration up to at least 50 pmol/μL. The measured spray current is approximately 30% of the calcualated current corresponding to a given concentration of quaternary ammonium ions in solution leaving the nebulizer at a given flow rate. The results obtained show that tetraakylammonium salt give the best sensitivitly with no influence of sample‐ion shape or counter ion. The dynamic range for full‐scan mass spectra of the compounds studied appears to the linear for concentrations from 0.1 pmol/μL (10−7M) up to 10 pmol/μL (10−5M).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.