Abstract

The response time of an optical-fibre pH sensor with a design discussed in this paper depends upon the matrix quantity and particle size, the indicator solution concentration, direction and magnitude of the pH change, shape of the sensor tip, temperature of the analyte solution and transition interval of the indicator relative to the pH of the solution. This paper reports studies of the response time of optical-fibre pH sensors based on the use of the indicators bromophenol blue and bromothymol blue measured using diffuse reflectance spectrometry.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.