Abstract

The qualitative changes in pyrolysis mass spectra obtained under various experimental conditions were studied by factor analysis. The parameters involved are the pyrolysis method (oven vs. filament), the method of cleaning the pyrolysis wire, the choice of the suspending liquid, the sample size, the filament position, the temperature-time profile and the temperature of the expansion chamber.For a number of these parameters, factor analysis of the data enabled a physico-chemical interpretation of the variations in the pyrolysis mass spectra to be made.An increased formation of dehydrogenated fragments was observed at higher equilibrium temperatures and for oven pyrolysis. Condensation of less volatile pyrolysis products at relatively cold parts of the inlet system (reaction tube and expansion chamber) was clearly observed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.