Abstract
We report the installation and performance evaluation of a probe aberration-corrected high-resolution JEOL JEM-ARM200F transmission electron microscope (TEM). We provide details on construction of the room that enables us to obtain scanning transmission electron microscope (STEM) data without any evident distortions/noise from the external environment. The microscope routinely delivers expected performance. We show that the highest STEM spatial resolution and energy resolution achieved with this microscope are 0.078 nm and 0.34 eV, respectively. We report a direct comparative evaluation of the performance of this microscope with a Schottky thermal field-emission gun versus a cold field-emission gun. Cold field-emission illumination improves spatial resolution of the high current probe for analytical spectroscopy, the TEM information limit, and the electron energy resolution compared to the Schottky thermal field-emission source.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.