Abstract

In this study, graphitic carbon nitride (g-C3N4) was used as a coating material for solid-phase microextraction (SPME) applications. Coupled to gas chromatography (GC), the extraction ability of the SPME fiber was investigated and compared with the commercial fibers of 100μm PDMS and 85μm CAR/PDMS using six target analytes including deltamethrin, nerolidol, amphetamine, dodecane, ametryn and acrylamide. The g-C3N4 coating revealed excellent extraction ability and durability comparing with those of the commercial fibers due to its loose structure and unique physicochemical properties. The repeatability for each single fiber was found to be 3.46% and reproducibility for fiber to fiber was 8.53%. The g-C3N4 SPME fiber was applied to the determination of acrylamide in potato chips, the linearity and detection limit was 0.5-250μgg−1 and 0.018μgg−1, respectively.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.