Abstract

Recently, EBT has been proven as the marker of airways inflammation like asthma and proposed as the non-invasive tool. Although EBT device has already been made but this device is rarely used for patients in Indonesia. The aim of this study is to develop a prototype which accurately measures EBT and is comfortably used by patients including children. This prototype was made using SHT11 as a sensor of EBT which is integrated on a thermal flask 0.5L. This flask filled up patient breath and the temperature of air breath was measured. The EBT of twelve healthy samples and seven samples with asthma was examined using this prototype, the measurement was done within three minutes for all of them. The test results of EBT on healthy samples obtained the median is 33.9°C within of 33.0°C - 34.7°C and EBT on asthma samples obtained median is 35.0°C within the range 34.9°C - 36.0°C.